Description:
DIGITALLY TUNED VARIABLE FILTER
Information
The Models 3340 and 3342, illustrated in Figure 1, are
variable, electronic Filters with digitally tuned cutoff
frequencies over the range from 0. 001 Hz to 99. 9 kHz.
The passband gain is unity (0 db) or 10 (20 db), with
attenuation rates of 48 db per octave (Model 3340) or 96 db
per octave (Model 3342} outside the passband. Maximum
attenuation is greater than 80 db, and the output hum and
noise is less than 500 microvolts. The Model 3340 can
function as either a High-Pass or Low-Pass Filter. In the
High-Pass mode of operation the maximum input signal isi
plus or minus 7 volts peak in the 0 db gain position, plus
or minus :0. 7 volts peak in the 20 db gain position, and
the upper 3 db point occurs at approximately 1 MHz. In the
Low-Pass mode the Filter is direct-coupled, and the combined
ac plus dc input signal should not exceed 7 volts peak in
the 0 db gain position and 0, 7 volts peak in the 20 db gain
position. The models 3341 and 3343 are identical to the 3340
and 3342, respectfully, except for the elimination of the
lowest band (. 001}. The Model 3342 consists of two of the
Model 3340 mounted in a single cabinet isolated from each
other with independent power supplies, and input and output
connectors. ;When these two Filter channels are switched to
the same mode of operation and connected in series with both
channels set to the same cutoff frequency, the Model 3342
will function as a High-Pass or Low-Pass Filter with an
attenuation rate of 96 db per octave. If the two channels
are connected in series, and one channel is operated in the
Low- Pass mode and the other channel in the High-Pass mode,
the Model 3342 will function as a Band-Pass Filter with an
attenuation rate of 48 db per octave outside the pass-band.
When the two channels are connected in parallel, as
described in Section 2. 3. 2. 3, the Model 3342 will
function as either a Band-Reject Filter with cutoff
frequency limits from 0. 001 Hz to 99. 9 kHz or provide a
null at any frequency between 0. 003 Hz and 30 kHz.
The Filter has a maximally flat (Butterworth} characteristic
when the FUNCTION switch(s} is in the Low-Pass MAX FLAT
position. For pulse-type wave-forms this switch should be in
the Low-Pass RC position, optimum for transient-free
filtering. The term (RC} as used in this manual may be
misleading. It is not an 8 pole RC Filter. A certain amount
of feedback still remains in 3 of the two-pole sections.
Manual Type:
Service and User Manual
Pages:
30
Size:
1.53 Mbytes (1601717 Bytes)
Language:
english
Revision:
Manual-ID:
Date:
Quality:
Scanned document, all readable.
Upload date:
2015 09 06
MD5:
d84906a7c5573d9571909ea3709f8d3a
Downloads:
1977
Information
GENERAL
DESCRIPTION ...
1
2 OPERATION.
...
8
3 CIRCUIT
DESCRIPTION ...
14
4
MAINTENANCE ... : ... ; ...
18
5 CALIBRATION AND
ADJUSTMENT ... 23