Informações:
                                    The Servo Ventilator 900 C/D/E consists of two sepa-­
rate units. 
The pneumatic unit comprises the gas conduction system with
two pressure transducers, two flow transducers and two servo
valves. Each transducer, continuously delivers its actual
value to the electronic unit. The servo valves are used as
CONTROLLING ­ELE­MENTS for the control of inspiratory and
expiratory gas conduction.

The electronic unit comprises three controlling systems, two
for regulation of inspiration and one for regulation of
expiration. The three controlling systems will be in use,
one at a time, depending on the TIMING control and the
se­lected ventilation MODE.

Estes manuais estão disponíveis para os equipamentos acima:

Tipo manual: Manual de serviço Maquet-10553-Manual-Page-1-Picture
Páginas: 55
Tamanho: 2.88 Mbytes (3022360 Bytes)
Língua: Inglês
Revisão:
Manual-ID/número:
Data:
Qualidade: Documento eletrônico, digitalização, bem legível.
Upload de data:
MD5: 65c3972cef97500d50e5ef52836b313a
Downloads: 23 desde 08 Março 2018
Informações:
The Servo Ventilator 900 was orginally manufactured and sold
by Siemens. Please note that this Service Manual has only
been re-labeled from Siemens to MAQUET. No further
significant changes are made. Thus, some information in the
Service Manual may be outdated.


Introduction . . . . . . . . . . . . . . . . . . 	 4
Principle diagram. . . . . . . . . . . . . . . . 	 6
Pneumatic unit
	 Gas flow through the ventilator . . . . . . . . . 
 	8
	 Gas supply. . . . . . . . . . . . . . . . . 
 	9
	 Bellows and spring assembly. . . . . . . . . .
 	10
	 Gas supply unit. . . . . . . . . . . . . . .
 	11
	 O2cell . . . . . . . . . . . . . . . . . . . 
 	12
	 Relief/safety valve and working
	 pressure manometer. . . . . . . . . . . . .
 1
 	 2
	 Flow transducer. . . . . . . . . . . . . . . 
 	14
	 Pressure transducer . . . . . . . . . . . . .
 	15
	 Inspiratory valve and step motor . . . . . . . . .
 	16
	 Expiratory valve and pull magnet . . . . . . . .
 	17
	 Non-return valve. . . . . . . . . . . . . . . 
 	17
Transducer interface. . . . . . . . . . . . . . .
 	18
Electronic unit
	
 Principle of operation. . . . . . . . . . . . . 
 	19
	
 Reference level generation. . . . . . . . . . . 
 2
 	 0
	
 Regulation of inspiration. . . . . . . . . . . .
 	23
	
 Timing. . . . . . . . . . . . . . . . . . . 
 	28
	
 Monitoring . . . . . . . . . . . . . . . . . 
 	33
	
 Regulation of expiration. . . . . . . . . . . . 
 	39
	
 Voltage supply . . . . . . . . . . . . . . . .
 	41
	
 Input/output. . . . . . . . . . . . . . . . .
 	42
Disassembling
	
 Disassembling the pneumatic unit . . . . . . . .
 4
 	 4
	 Disassembling the electronic unit.. . . . . . . . 
 	44
	 Replacement of step motor assembly
	 and pull magnet. . . . . . . . . . . . . . . 
 	45
	 Replacement of power supply unit . . . . . . . .
 4
 	 5 	
	 Replacement of safety catches . . . . . . . . . 
 	45
	 Replacement of flow transducer . . . . . . . . .
 	46
Adjustments
	 Adjustment of meters Z1 and Z2. . . . . . . . .
 	 4
6
	 Adjustment of valve arm. . . . . . . . . . . .
 	46
	 Replacement and adjustment of R5 . . . . . . .
 	47
	 Removing the front panel knobs . . . . . . . . .
 	47
	 Front panel potentiometer adjustments. . . . . . .
 	48
Troubleshooting. . . . . . . . . . . . . . . .
 	49
Index. . . . . . . . . . . . . . . . . . . . .
 	52
Block diagram. . . . . . . . . . . . . . . . . 
 	56