Equipment information
Associated Research Incorporated 4045AI Level Meter
Manufacturer:
Image 1 of 1
If you have any other photos or manuals for the
Associated Research Incorporated 4045AI you can upload the files here .
Model:
4045AI
Date:
1987
Category:
Group:
Description:
Megohmmeter
Information
1. Wide, easily read 4-1/2 inch AC/DC kilovoltmeter with 2%
of full scale accuracy.
2. Detachable 5 foot return (ground) lead. This lead is
always used in performing tests and is grounded for safety.
3. Detachable 5 foot high voltage lead with flexible rubber
insulation for easy handling. Control panel Jacks are well
recessed for safety.
4. Lighted AC and DC indicators on the control panel which
show that the High Voltage circuit Is enabled ("Power 0a")
and Indicate which node
of operation was selected by rotating the power switch to
either the AC or DC position*
5. Lighted Breakdown and Leakage Indication. If excessive
current is drawn during the Hypot* test, the yellow leakage
indicator will illuminate and an audio alarm will sound.
Should any arcing occur during the Hypot* test, both the
yellow leakage indicator and the orange breakdown indicators
will illuminate. To reset the unit, simply decrease the
output voltage to zero, place the power switch Into the
"Off" position and remove whatever is causing the failure
from the output leads.
6. Voltage adjust knob. This control varies the High voltage
from near zero to the full 5000 volts AC or 5000 volts DC.
The voltage can be adjusted only when the high voltage
circuit is energised. It is recommended that the voltage
control be returned to zero at the end of each test.
7. Detachable 7-1/2 fooc power cord with standard coanector.
8. Line fuse accessible from the rear. "Shock, safe" fuseholder.
9. Separate AC and DC Current Trip (sensitivity)
adjustments^ The current at which the Failure alarm will
trip is adjustable from approximately .3 mA to 5 mA AC or 3
mA DC. Turn the controls counterclockwise to decrease the
current setting and increase sensitivity.
Turn them clockwise to increase the current setting and
decrease sensitivity.
Manual Type:
User Manual
Pages:
13
Size:
825.90 Kbytes (845724 Bytes)
Language:
english
Revision:
Manual-ID:
Date:
1987 01 01
Quality:
Scanned document, reading partly badly, partly not readable.
Upload date:
2017 10 21
MD5:
987e1ad9da1a18996ea03dea1203041d
Downloads:
410
Information
This manual discusses the reasons why dielectric withstand
Tests are required, The precautions necessary to ensure that
The tests are done safely, and how such Tests are
accomplished on various classes of materials, components and
products.
In Later chapters, the features of this Associated Research
Model Hypot® are explained, along which a step by step guide
Co its proper use. Finally, technical data is presented.